3.5 Operation screen reference
3
-
51
FRA 51602
3.5.3 Sweep dialog
■ [SEQ SWEEP]
: Set sequence measurement.
This function performs measurements in the order set in the setting memory of the instrument. It is
possible to do measurements at different signal levels for each frequency range. For more information in
sequence measurements, refer to "
5.11 Sequence measurement
".
○ Setting method
< List operations (confirming) >
• OFF, 1 to 20
: No sequence measurements, or 1 to 20
■ [SWEEP LOWER] / [SWEEP UPPER]
: Set the upper and lower sweep frequencies.
This sets the frequency range for performing sweep measurements.
○ Setting method
< Numerical value settings (immediate setting) >, < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
The setting range is from 10 µHz to 2 MHz and the resolution is 10 µHz. The lower frequency limit
must be less than the upper frequency limit.
○ Function key functions
• [MKR -> SWEEP]
: Set the frequency of the main marker and the delta marker to the range
of the sweep frequency.
• [GRAPH -> SWEEP]
: Set the sweep frequency range to the range of the X axis (frequency
range) of the graph.
• [MKR -> FREQ]
: Copy the main marker frequency to the sweep lower frequency limit
(for
[SWEEP LOWER]
) or to the upper frequency limit (for
[SWEEP
UPPER]
).
• [DIRECT NEW]
: Set the frequency with the numerical input dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the frequency setting with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [uHz]
: Set with the entered value in units of µHz.
• [mHz]
: Set with the entered value in units of mHz.
• [Hz]
: Set with the entered value in units of Hz.
• [kHz]
: Set with the entered value in units of kHz.
• [MHz]
: Set with the entered value in units of MHz.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...