5.4 Oscillator control
5
-
9
FRA 51602
5.4.2 Oscillator stop mode
You can specify the stop (OFF) timing of oscillator output by selecting from
[QUICK]
and
[0° SYNC]
. The
setting is done using
【
OSC
】
–
[STOP MODE]
.
[QUICK]
: The output is turned off immediately, regardless of the signal phase.
[0° SYNC]
: The oscillator output is not turned off until it reaches 0° phase.
Figure 5-3
Oscillation stop mode setting and change in output voltage
The "0° SYNC"
setting is used for measuring electrochemical impedance (EIS) and other such
measurements. The oscillator always starts at 0° phase when turned on, so the "0° SYNC
"
setting means
that the average voltage over the time from on to off (DC component) is zero and the total amount of
electricity injected to the DUT is the same after the measurement as before the measurement.
Please note that it may take about one cycle before the output actually ceases, depending on the timing of
turning the output off. (For 10 µHz, it will take 10
5
s
≒
28 hours to turn off in the worst case.)
Please also note that the actual oscillation stop phase may deviate from 0° if the oscillation frequency
exceeds about 100 kHz.
When the oscillator stop mode is set to "0° SYNC
"
, the SLOW setting of the oscillator on/off mode
is ignored. ("0° SYNC
"
setting of the oscillator stop mode takes priority.) The oscillator is turned off as
soon as the 0° phase timing is detected.
0° SYNC
QUICK
AC OFF
Oscillator stop mode
Time
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...