9.1 Oscillator section
9
-
2
FRA 51602
Unless stated specifically otherwise, the assumed conditions are sinusoidal waveform, a temperature of
23°C ±5°C, no more than 12 hours has passed since calibration has been performed after a warm-up period
of at least 30 minutes, and change in ambient temperature of not greater than ±5°C after calibration.
Unless stated specifically otherwise, the shell (outer cover) and signal grounds are connected to the
instrument enclosure (not isolated).
9.1 Oscillator section
The oscillator section can output an AC signal and the DC bias for measurement. The AC signal is output
from the
OSC
connector on the front panel. The DC bias can be output from either the
OSC
connector on
the front panel or the
DC BIAS OUT
connector on the rear panel. You can select the connector from which
the DC bias is output. If the
OSC
connector on the front panel is selected for DC bias output, the signal
output from the
OSC
connector is a composite of the AC signal and the DC bias.
9.1.1 Front panel OSC
Connector
Insulated BNC connector (front panel,
OSC
)
Output waveform
Sinusoidal, square, or triangular
Frequency
Setting range
10 µHz to 2 MHz
The quality of square and triangular waveforms deteriorates at frequencies
above 100 kHz.
Setting resolution
10 µHz
Accuracy
±10 ppm
Conditions: The instrument is operating on the internal reference clock.
AC signal amplitude
Setting range
0 to 10 Vpk Conditions: No load
Setting resolution
3 digits or 0.01 mVpk, whichever is greater
Accuracy
±0.3 dB
Conditions: sinusoidal waveform, 1 kHz, 100 mVpk to 10 Vpk, no load
Frequency characteristic ±0.3 dB or less (100 kHz or less)
±1 dB or less (1 MHz or less)
±3 dB or less (2 MHz or less)
Conditions: 1 kHz reference, sinusoidal waveform, 100 mVpk to 10 Vpk, 50
Ω load
Distortion (sinusoidal) 0.2% or less, supplementary value
Conditions: 100 kHz or less, BW 500 kHz, 10 Vpk output
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...