3.5 Operation screen reference
3
-
44
FRA 51602
■ [10MHz REF]
–
[IN]
: Set the reference clock input.
This is used to obtain higher frequency accuracy and stability by using an external reference clock that is
more accurate than the internal reference clock of the instrument. This is also used to make frequency
accuracy common among multiple instruments.
○ Setting method
< List operations (confirming) >
• DISABLE
: Use the internal reference clock.
• ENABLE
: Use an external reference clock if an external reference clock is available.
Even if
ENABLE
is set, the internal reference clock of the instrument will be used if an external reference
clock is not detected or if the frequency differs from the specified frequency. The reference clock on
which the instrument is currently operating is indicated by the status display in the upper part of the LCD
screen.
■ [10 MHz REF]
–
[OUT]
: Set the reference clock output
This is set when the reference clock on which the instrument is operating is output from the
10 MHz
REF OUT
connector on the rear panel. This is done so that this instrument and other devices can operate
with the same frequency accuracy.
○ Setting method
< List operations (confirming) >
• DISABLE
: The reference clock is not output.
• ENABLE
: The reference clock is output.
When the instrument is operating on an external reference clock, the output frequency is the same as that
reference clock.
■ [AMPLITUDE COMPRESSION]
: Set the amplitude compression function
This is used to automatically adjust the output amplitude of the oscillator to supply a constant amplitude
to the circuit being measured. For more information on the amplitude compression function, refer to "
5.7
Amplitude compression
".
INT : Operating on the internal reference clock
EXT : Operating on the external reference clock
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...