9.2 Analysis input section
9
-
5
FRA 51602
Capacitance relative to enclosure
200 pF or less
Frequency range
10 µHz to 2 MHz
Maximum measurement voltage
600 Vrms
When the provided signal cable is used.
When the other signal cables are used, the voltage is restricted to the rated
voltage of the cable or to 600 Vrms, whichever is less.
Over-level detection
Setting range
0 to 600 Vrms
Setting resolution
3 digits
Processing
Over lamp lights.
Buzzer warning sound (can be turned on / off)
Stop sweep measurement (can be turned enable / disable)
Over-level detection occurs when the amplitude or maximum measurement input voltage that is set as
described above is exceeded. (See "
9.2 Analysis input section
, ● Measurement range”.)
Dynamic range
140 dB (10 Hz to 1 MHz)
80 dB (1 MHz to 2 MHz)
The maximum measurable gain, limited by the noise floor and cross talk
between channels.
Conditions: The larger channel input is at least 10 Vpk, integration is 300
cycles, and the measurement range is AUTO for both channels.
Input weighting
Setting range is 0 to 1.12, resolution 6 digits or 1E-12
CH1 and CH2 are set independently
The weight is a coefficient that is applied to measurement results of CH1 or
CH2 to correct the attenuation or gain of probes, amplifiers, or other such
equipment that is connected to the instrument.
Phase inversion function ON or OFF
This function is used when the polarity of connection to the circuit being
tested is reversed.
The isolation specifications for between the oscillator (
OSC
) and analysis input (
CH1
or
CH2
) and the
instrument enclosure is illustrated in
Figure 9-1
. The isolation specifications between the
OSC
,
CH1
, and
CH2
signal connectors is illustrated in
Figure 9-2
.
Figure 9-1
Isolation voltage specifications relative to the enclosure
OSC
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CATII
300 V CAT
Ⅲ
CH1
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
CH2
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...