8.5 Performance test
8-4
FRA 51602
8.5 Performance test
8.5.1 Introduction
The performance test is a part of preventive maintenance and is performed to prevent degradation of
instrument performance. The testing is done for the acceptance inspection, periodic inspections, and
whenever it is necessary to check performance, such as after repair. This section describes a simple test that
can be done with commercially available measuring instruments.
More advanced testing is available at charge and can be requested of the NF Corporation or its authorized
agent.
If the performance test indicates that the instrument does not satisfy the specifications, calibration or repair
is required. In that case, please contact the NF Corporation or its authorized agent.
1) The test environment
The performance test should be performed under the following conditions.
Ambient temperature
23 ±5°C
Ambient humidity
20 to 70% RH, no condensation
Power supply voltage
100 to 230 VAC ±10%, however 250 V or less
Warm-up time
30 minutes or more
2) Preparation for testing
Before doing the test, initialize the settings and calibrate the instrument. To initialize the instrument,
select
【
Other
】
–
[RESET]
from the menus. For more information, refer to "
■ Initialization
" in "
4.2.1
Common operations
". To calibrate the instrument, select
【
Calib.
】
–
[CALIBRATION]
from the menus.
For more information, refer to "
5.9.1 Calibration
".
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...