3.5 Operation screen reference
3
-
43
FRA 51602
■ [STOP MODE]
: Set the oscillation stop phase.
This selects whether to stop the oscillation instantaneously or to wait until the 0° phase of a sine function
before stopping the oscillation. This is useful for the measurement of electrochemical impedance because
the amount of charge transfer can be made zero by equalizing the areas of the positive and negative sides
of an AC sinusoidal signal.
○ Setting method
< List operations (confirming) >
• QUICK
: Turn off instantly.
• 0° SYNC
: Turn off after waiting for 0° phase (stop oscillation).
When
0° SYNC
is set, it takes up to one cycle of the signal at the longest from the time the output off
operation is performed to the time the output is actually turned off. Please note that, at 10 µHz, it takes up
to 10
5
seconds (27.8 hours) at the longest.
If the
[ON/OFF MODE]
is set to
SLOW
and the
[STOP MODE]
is set to
0° SYNC
, the
[STOP
MODE]
setting has priority.
■ [WAVEFORM]
: Select the waveform.
This selects the oscillator output waveform from among sinusoidal, square, and triangular. Normally,
sinusoidal is used. Main characteristics such as amplitude accuracy and measurement accuracy are
specified only for the sinusoidal waveform.
○ Setting method
< List operations (confirming) >
• SINE
:
Sinusoidal
• SQR
: Square
• TR I
: Triangular
The square and triangular waveforms should be used for frequencies of 100 kHz and below. At higher in
frequencies, waveform quality decreases.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...