8.5 Performance test
8-7
FRA 51602
8.5.3 Oscillator AC amplitude accuracy
This instrument
: Set the frequency to 1 kHz.
Measuring instrument
: Set the digital multimeter to AC voltage mode (true RMS).
Connections
: Connect the
OUTPUT OSC
connector of this instrument to the digital
multimeter with a coaxial cable.
Measurement method
: Set the AC amplitude of this instrument to 100 mVpk, 1 Vpk, and 10
Vpk, and read the AC voltage from the multimeter.
The amplitude accuracy is calculated from the measured AC voltage with
the formula shown below.
01
.
3
(Vpk)
amplitude
AC
Set
(Vrms)
voltage
AC
M easured
log
20
(dB)
accuracy
Amplitude
10
Judgement
AC amplitude
setting
Measured AC
voltage
Amplitude accuracy
Passing range
100 mVpk
_ _ _. _ mVrms
_ _._ _ dB
-0.30 to +0.30 dB
1 Vpk
_. _ _ _ Vrms
_ _._ _ dB
-0.30 to +0.30 dB
10 Vpk
_ _. _ _ Vrms
_ _._ _ dB
-0.30 to +0.30 dB
FRA 51602
OSC CH1
CH2
Digital multimeter
BNC-BNC
coaxial cable
BNC-banana
adapter
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...