3.5 Operation screen reference
3
-
52
FRA 51602
■ [SWEEP RESOLUTION]
: Set the sweep resolution.
This sets the resolution of the sweep frequency. The number of data points obtained in a sweep
measurement is the resolution value set here +1.
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
The setting range is from 3 to 20,000 and the resolution is 1. It is possible to select between linear
sweep (common difference) and log sweep (common ratio).
○ Function key functions
• [DIRECT NEW]
: Set the frequency with the numerical input dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the frequency setting with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [Log steps/sweep]
: Set the entered value as the log sweep resolution.
• [Lin steps/sweep]
: Set the entered value as the linear sweep resolution.
■ [REPEAT]
: Set the repeat mode.
If repeat is on, the measurement is repeated. For sweep measurement, the sweep measurement is started
again at the initial frequency. The measurement is repeated until the
STOP
key is pressed. If repeat is off,
the measurement is performed one time and then stopped.
○ Setting method
< List operations (confirming) >
• OFF
: Repeat off
• ON
: Repeat on
■ [FREQUENCY CHANGE MODE]
: Set the frequency change phase.
This selects whether to change the frequency immediately or to wait until the 0° phase of a sine function
to change the frequency. When performing electrochemical impedance measurements, etc., the area of the
plus side and the area of the minus side of an AC sinusoidal signal are the same, so the total charge
transfer to the DUT can be made to be zero.
○ Setting method
< List operations (confirming) >
• ASYNC
: The frequency is changed instantaneously.
• SYNC
: The frequency is not changed until the 0° phase.
When the setting is
SYNC
, it takes up to one cycle of the signal at the longest until the frequency is
changed (including when a sweep is in progress). Please note that, the maximum length takes up to
10
5
seconds (27.8 hours) at the longest for a frequency of 10 µHz.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...