5.10 Differentiation and integration function
5
-
39
FRA 51602
5.10 Differentiation and integration function
This function performs differentiation, second-order differentiation, integration, and double integration of
the measurement data in the time domain. This is used when doing displacement–speed–acceleration
conversion in the measurement of transfer characteristics (gain-phase measurement).
Figure 5-20
Conversion among displacement, speed, and acceleration
by differentiation and integration
Differentiation in the time domain corresponds to multiplying the measurement data in the frequency
domain by jω and the second derivative corresponds to multiplication by (jω)
2
. Similarly, integration
corresponds to dividing the frequency domain measurement data by jω and the double integral corresponds
to division by (jω)
2
. The phase is changed by +90° by differentiation and by −90° by integration.
The setting is done by selecting
【
Measure
】
–
[d/dt]
, pressing the
ENTER
key, and then selecting the
calculation item. This function is set to
[OFF]
if integration and differentiation are not to be done.
■ Cautions concerning differentiation and integration
If differentiation and integration is enabled, the amplitude of CH2 is normalized to 1.0 (Vrms) for display.
The gain (CH1/CH2) and phase values can be used to produce corrected results, but please note that the
CH1 amplitude and the CH2 amplitude that are displayed differ from the measured amplitudes.
Displacement
Speed
Acceleration
Differentiation
Differentiation
Integration
Integration
Second-order differentiation
Double integration
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...