4.2 Simple operation for the first use
4
-
13
FRA 51602
If the OSC output of the instrument is amplified with a power amplifier, it is possible to measure the
impedance by applying a large voltage/current to the DUT. A large amplitude or DC bias (voltage/current)
can be used to measure a change in the impedance characteristic caused by a nonlinearity such as saturation
of a magnetic or dielectric material.
Figure 4-7
Example of connections for measuring impedance with high-power
1
~
2
Zx
Current detection
resister
Rs
i
i x Rs
v
Power amplifier