3.5 Operation screen reference
3
-
48
FRA 51602
3.5.2 Measure dialog
■ [INTEGRATION MODE]
: Set the integration mode.
This sets the integration mode. For more information on integration operation, refer to section "
4.4
Integration
".
○ Setting method
< List operations (confirming) >
• FIX
: Set manual integration.
• SHORT
: Set automatic integration (short time, high variance).
• MED
: Set automatic integration (medium time, medium variance).
• LONG
: Set automatic integration (long time, low variance).
■ [INTEGRATION]
: Set the integration count or time.
This sets the number of times (or the duration) for manual integration. Increasing this value results in
lower noise and cleaner measurement results, but longer measurement time. For information on the
integration operation, refer to "
4.4 Integration
".
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
For number of times, the setting range is 1 to 9,999 cycles and the resolution is one cycle. For duration,
the setting is 0 to 9,990 seconds and the resolution is 3 digits or 0.1 ms, whichever is greater.
Integration is always performed for at least one cycle of the signal, even if the set value is 0 seconds.
○ Function key functions
• [DIRECT NEW]
:
Set integration with the numerical input dialog
【
NUM
】
.
• [DIRECT EDIT]
:
Edit integration with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [CYCLE]
: Set the entered value in units of cycle.
• [msec]
: Set the entered value in units of milliseconds.
• [sec]
: Set the entered value in units of seconds.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...