7.1 Error messages
7
-
4
FRA 51602
Table 7-2
Panel operation errors (continued)
No.
Error message
Description and cause
Handling
-6145 There is already a file of
the same name.
A file of the same name already exists,
so the specified name cannot be used to
rename the file.
Specify a different file
name.
-6146 USB storage overflow.
There isn't sufficient available space on
the USB memory device.
Insert a USB memory
device that has sufficient
available space.
-6147 File load error due to
illegal header.
The file could not be loaded because it
is an invalid file (invalid header).
Specify a valid file.
-6148 File load error due to
illegal data.
The file could not be loaded because it
contains invalid data.
Specify a correctly
formatted file.
-6149 Failed to save.
The process of saving the file to the
USB memory device failed.
Remove the read-only
attribute from the file.
-6150 Failed to load.
The process of reading the file from the
USB memory device failed.
Specify an undamaged
file.
-6151 Failed to delete.
The process of deleting the file from
the USB memory device failed.
Remove the read-only
attribute from the file.
-6152 Can't load empty
memory.
The process of reading the
measurement data failed because the
internal memory is empty.
Specify in internal
memory that is not
empty.
7.1.3 Errors that occur during measurement
The errors that may occur during measurement are listed in
Table 7-3
.
Table 7-3
Errors that occur during measurement
No.
Error message
Description and cause
Handling
-3072 Measurement terminated
due to measurement
points over 20,001.
The number of measurement points
exceeded the limit of 20,001, so
measurement was stopped.
When performing sequence
measurement, reduce the
number of measurement
points at each step.
When performing automatic
high density sweep
measurements, change the
threshold setting.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...