background image

USER MANUAL 

 

MI 806 

pH/EC/TDS/Temperature 

Portable Meter 

 

 

 

 

 

 

GENERAL OPERATION 
 

• The meter is supplied 
complete with a 9V battery. 
Remove the battery cover 
on the back of the meter 
and install the battery while 
paying attention to its 
polarity. 
• Connect the 

MA851D/1 

probe to the meter and turn 
it on by pressing the 
ON/OFF key. 
• At start-up, the LCD shows the 
percentage of the remaining battery life 
for a few seconds, and then the current 
measurement. 
• To select the temperature 
measurement unit (°C or °F), press and 
hold the “ON/OFF” key until “TEMP” and 
the current temperature unit are 
displayed on the secondary LCD. Use 
the “SET” button to select the unit, then 
press “ON/OFF” a couple of times to 
return to normal mode. 
• To activate the HOLD 
function, keep pressed 
the “HOLD” key. The 
measured value will be 
frozen on the display 
and the “HOLD” 
message appears on 
the secondary LCD. 
• To set the measurement range (pH, 

μS, ppm), press the “RANGE” key. The 
chosen mode will be stored until next 
change. 
• Before taking any measurement, make 
sure that the meter has been calibrated 
(the “CAL” tag is displayed on the left 
lower corner of the LCD). 
• After measurements, switch the meter 
off by pressing the “ON/OFF” key. The 

“OFF” message appears on the LCD; 
release the button. 
• Store the electrode with a few drops of 
storage solution in the protective cap. 
 

TAKING pH MEASUREMENTS 

• Always remove the electrode 
protective cap before taking any 
measurement. If the electrode has been 
left dry, soak the tip (bottom 2.5 cm) in 

M10000 

rinse solution for a few minutes 

to reactivate it. 
• Select the pH range with the “RANGE” 
key. 
• Immerse the tip (2.5 cm) of the probe 
into the sample and stir gently. 
• Read the pH value when the clock 
symbol stops blinking. 

 
pH CALIBRATION PROCEDURE 

A)  Preparation, Buffer solutions: 
1. 

pH 7.01 

6.86 (MA9007 / MA9006) 

2. 

pH 4.01 

(

MA9004

) for acidic 

calibration (pH <  7) or 

pH 10.01 / 9.18 

(

MA9010 / MA9009

) for alkaline range 

(pH > 7). 
Use two beakers for each buffer 
solution: one beaker for rinsing the 
probe, the other for calibration. In this 
way the cross contamination between 
solutions is minimized. 
B) Procedure: 
• Enter the pH mode with the “RANGE” 
key. 
Select calibration buffer set:  press and 
hold the “ON/OFF” key until the LCD 
shows “TEMP”. 
Press again this key and the “BUFF” 
message will appear; then select the 
desired buffer set with the “SET” key: 
“7.01 pH BUFF” (for standard solutions: 
pH 4.01, 7.01, 10.01) or “6.86 pH BUFF” 

(for NIST solutions: pH 4.01, 6.86, 9.18). 
Press “ON/OFF” again to exit. 
• Remove the protective cap, then 
immerse the probe in the first buffer 
solution. 
• Press and hold the 
“ON/OFF” key until 
“CAL” is displayed on 
the lower LCD. Release 
the button and the 
message “7.01 pH USE” 
(or “6.86 pH USE” for 
NIST buffer set) will be 
displayed. 
• For a single point pH calibration, place 
the probe in any buffer from the selected 
set (e.g. pH 4.01 or 7.01/6.86 or 
10.01/9.18). The meter will automatically 
recognize the buffer value, store the 
calibration point and return to normal 
mode. 
• If using pH 7.01 (or pH 6.86), after 
recognition of the buffer press the 
“ON/OFF” button to return to normal 
mode. 
• For a two point pH calibration, place 
the probe in pH 7.01 buffer (or pH 6.86, 
if the NIST buffer set was selected). 
The meter will recognize 
the buffer value and 
then display pH 4.01 
USE. 
• Place the probe in the second buffer 
(pH 4.01 or 10.01, or, if using NIST, pH 
4.01 or 9.18). When the second buffer is 
recognized, the LCD will display “OK” for 
1 second and the meter will return to 
normal mode. 
• The “CAL” tag is displayed on the LCD 
to indicate that the meter is calibrated. 

Reviews: