5.11 Sequence measurement
5
-
41
FRA 51602
■ Cautions concerning sequence measurement
• The validity check is not performed for the sweep frequency range of each setting memory number.
Depending on the settings in the memory, the sweep frequencies may be greatly separated or the sweep
ranges may overlap when the memory number is changed. Nevertheless, the measurements in sequence
measurement are performed in the order of the setting memory number.
• For marker operations on sequence measurement data, the markers are moved in the order in which the
sequence measurements were performed. If the high and low values of the sweep range are reversed
between setting memories, turning the control knob to the right (clockwise) may move the marker in the
direction of lower frequencies.
• The total number of sweep frequency points that can be measured in a single sequence is the sum of the
number for memory 1 through memory N (largest memory number) and the maximum total is 20,001
frequency points. If the maximum number is reached while sequence measurement is still in progress,
the measurement is stopped at that time. No check is made for the limit on the number of sweep points
before the measurement begins. If the automatic high density sweep function (SLOW sweep) is used,
please note that there may be an unexpected increase in the number of frequency points.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...