9.4 Analysis processing section
9
-
13
FRA 51602
Measurement data items and display range
The actual measured values and display range for each parameter are constrained by the measurement
range and the frequency, etc.
dBR (gain dB) ±999.999 dB, resolution 0.001 dB
θ (phase)
±180°
-180.000° to 179.999°, resolution 0.001°
0 to 360°
0.000° to 359.999°, resolution 0.001°
-360 to 0°
-360.000° to -0.001°, resolution 0.001°
UNWRAP
-9999.999° to + 9999.999°, resolution 0.001°
GD (group delay)
±(1 a to 999.999 G)s or 0 s, resolution 6 digits or 1 as
(when the analysis mode setting is CH1/CH2 or CH2/CH1)
R (absolute gain)
0 to 999.999 G (ratio, unitless number), resolution 6 digits or 1 a
a (real part of gain)
±(1 a to 999.999 G) or 0 (ratio, unitless number), resolution 6 digits or 1 a
b (imaginary part of gain)
±(1 a to 999.999 G) or 0 (ratio, unitless number), resolution 6 digits or 1 a
(when the analysis mode setting is CH1 or CH2)
R (amplitude)
0 Vrms to 999.999 GVrms, resolution 6 digits or 1 aVrms
a (real part of amplitude)
±(1 a to 999.999 G) Vrms or 0 Vrms, resolution 6 digits or 1 aVrms
b (imaginary part of amplitude)
±(1 a to 999.999 G) Vrms or 0 Vrms, resolution 6 digits or 1 aVrms
Measurement accuracy for other than R and θ
Accuracy of measurement items other than R (absolute gain) and θ (phase). Where defined the accuracy
of R as Ra %, the accuracy of θ as θa, and the measurement value of θ as θx, specified in "
9.4 Analysis
processing section
, ● Measurement accuracy (fixed range) and ● Measurement accuracy (AUTO
range)".
GD (group delay)
±{θa / (APT × 360)} s
APT: The unit for aperture in the group delay display is hertz
a (real part of gain)
±(Ra / cos θx)%
b (imaginary part of gain) ±(Ra / sin θx)%
Error correction function
Equalizing
This function obtains the characteristics for the measurement target alone by
measuring the frequency characteristics of the measurement system (sensors,
cables, etc.) in advance and then eliminate that error component for the
measurement system when the actual measurements are performed later.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...