8.5 Performance test
8-10
FRA 51602
8.5.6 DC bias accuracy (DC BIAS)
This instrument
: Set the DC bias output to the
DC BIAS
connector on the rear panel.
Measuring instrument
: Set the digital multimeter to DC voltage mode.
Connections
: Connect the
DC BIAS
connector of this instrument to the digital
multimeter with a coaxial cable.
Measurement method
: Set the DC bias of this instrument to −10 V, 0 V, and +10 V, and read the
DC voltage from the multimeter.
Judgement
DC bias setting
DC voltage value
Passing range
−10 V
−_ _. _ _ _ V
−10.130 to −9.870 V
0 V
_ _ _. _ mV
−30.0 to +30.0 mV
+10 V
+_ _. _ _ _ V
+9.870 to +10.130 V
Digital multimeter
FRA 51602 rear panel
DC BIAS
BNC-BNC
coaxial cable
BNC-banana
adaptor
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...