9.2 Analysis input section
9
-
4
FRA 51602
9.1.2 Rear panel DC BIAS OUT
The DC bias is output only when the
DC BIAS OUT
is set as the output connector for the DC bias.
Connector
BNC connector (rear panel,
DC BIAS OUT
)
Setting range
-10 V to 10 V Conditions: no load
Setting resolution
10 mV
Accuracy
±(|1% of DC bias setting | + 30 mV)
Output resistance
600 Ω ±2%
9.2 Analysis input section
Input channels
2
Connectors
Insulated BNC connector (front panel,
CH1
and
CH2
)
Input impedance
1 MΩ ±2%, 20 pF ±5pF
Measurement range
30 mV to 600 V (rms), and AUTO. CH1 and CH2 can be set independentry.
Measurement range and maximum measurement input voltage
Measurement range
(rms)
30 mV
100 mV
300 mV
1 V
3 V
Maximum
measurement input
voltage
±42.5 mV
±142 mV
±425 mV
±1.42 V
±4.25 V
Measurement range
(rms)
10 V
30 V
100 V
300 V
600 V
AUTO
Maximum
measurement input
voltage
±14.2 V
±42.5 V
±142 V
±425 V
±849 V
There is restricted by frequency. See "● Maximum input voltage" below.
Maximum input voltage 600 V CAT II or 300 V CAT III (BNC signal to BNC ground)
Below 45 Hz:
600 Vpk (AC + DC)
45 Hz to 100 kHz:
600 Vrms (AC)
100 kHz to 2 MHz:
600 Vrms (AC) × 100 ÷ F
where F is frequency in units of kHz.
At 45 Hz and above, the voltage including DC must be 600 Vrms or less and
849 Vpk or less.
IMRR (isolation mode rejection ratio)
120 dB or more (DC to 60 Hz)
Conditions: Input shorted, integration 200 cycles or more, and common mode
voltage 10 Vpk
Isolation
600 V CAT II or 300 V CAT III (BNC ground to enclosure)
Below 45 Hz:
600 Vpk (AC + DC)
45 Hz to 100 kHz:
600 Vrms (AC)
100 kHz to 2 MHz:
600 Vrms (AC) × 100 ÷ F
where F is frequency in units of kHz.
At 45 Hz and above, the voltage including DC must be 600 Vrms or less and
849 Vpk or less.
Summary of Contents for FRA51602
Page 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Page 2: ......
Page 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Page 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Page 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Page 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Page 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Page 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Page 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Page 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Page 280: ...XVI FRA51602 Blank...
Page 282: ......
Page 284: ......
Page 285: ......
Page 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...