Acoustic Output Reporting Table (60601-2-37) C-19
Transducer Model: 35C50EA
Imaging Mode: PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.87
0.24
1.02
1.57
Index component value
0.16
0.24
0.66
1.02
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
1.38
P
(mW)
33.22
34.43
54.04
P
1×1
(mW)
13.21
25.43
z
s
(cm)
2.89
z
b
(cm)
2.23
z
MI
(cm)
0.50
z
pii
,
α
(cm)
0.50
f
awf
(MHz)
2.52
2.54
2.54
2.53
Other
Informatio
n
prr
(Hz)
699.00
srr
(Hz)
/
n
pps
/
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
82.24
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
85.56
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
114.61
p
r
at
z
pii
(MPa)
1.25
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
14.8cm
14.8cm
14.8cm
4.6cm
SV Position
6.0cm
15.0cm
6.0cm
2.0cm
Working Frequency
2.5
2.5
2.5
2.5
PRF
699
4399
10000
20000
SV
0.5mm
0.5mm
0.5mm
0.5mm
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...