7-4 Measurement
7.4
Measurement Accuracy
Table 1 Error of 2D Images
Parameter
Value Range
Error
Distance
Full screen
Within ±3%; or when the measured value is less
than 40 mm, the error is less than1.5 mm.
Area (Trace)
Full screen
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Area (ellipse,
circle)
Full screen
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Angle
Full screen
Within ±3%.
Circumference
Full screen
Within ±10%.
Distance (iScape) Full screen
Micro-convex probe: within ±15%
Other probe: within ±10%
Table 2 Volume Measurements
Parameter
Value
Range
Error
Volume
Full screen
Within ±10 %; or when the measured value is less than 64 cm
3
,
the error is less than 6.4 cm
3
.
Table 3 Time/Motion Measurements
Parameter
Value Range
Error
Distance
Full screen
Within ±3 %; or when the measured value is less
than 40mm, the error is less than 1.5mm.
Time
Timeline Display
Within ±2%.
Heart rate
Timeline Display
Within ±4%.
PW velocity
10 cm/s ~200cm/s
When the correction angle is
≤60º, the PW
velocity ≤±10%.
NOTE:
Within the selected field range, the measurement accuracy is ensured within the
range mentioned above. The accuracy specifications are performance in the worst
conditions, or based on the real test for the system, regardless of acoustic speed
error.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...