Acoustic Output Reporting Table (60601-2-37) C-31
Transducer Model: 65EB10EA_T
Imaging Mode: B-mode /Smart3D/Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.83
0.04
0.04
0.04
Index component value
0.04
0.04
0.04
0.02
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
2.04
P
(mW)
4.00
2.00
2.00
P
1×1
(mW)
1.79
0.89
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.65
z
pii
,
α
(cm)
1.65
f
awf
(MHz)
6.04
5.14
5.04
5.04
Other
Informatio
n
prr
(Hz)
5,096.0
0
srr
(Hz)
41.00
n
pps
1.33
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
158.75
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
4.50
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
9.36
p
r
at
z
pii
(MPa)
2.75
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
6.5cm
6.5cm
6.5cm
Focus position
2.0cm
0.5cm
0.5cm
0.5cm
Working Frequency
6.5
5
5
5
PRF
5096
8992
8992
8992
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...