12-30 Probes and Biopsy
NGB-010 metal/needle detachable needle-guided bracket
(1) Put on the transducer cover.
(2) Hold the transducer by one hand, select proper needle-guided bracket, and hold it
with the other hand, and align the grooves of the needle-guided bracket with the
tabs of the transducer, then push the needle-guided bracket forward, making the
grooves of the needle-guided bracket to match with the tabs of the transducer. Set
the needle-guided bracket at the desired position, turn tightly the knob of fixing
needle-guided bracket to fix the needle-guided bracket.
(3) Check manually to confirm the needle-guided bracket is securely installed on the
transducer.
(4) Select a proper guiding block and push it into the groove above the support of
needle-guided bracket, then turn tightly the knob of fixing the guiding block to fix the
guiding block on the support of needle-guided bracket.
(5) Insert a biopsy needle with the same specification as that of the guiding block into
the hole of the guiding block.