C-20 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 35C50EA
Imaging Mode: B+PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.92
0.48
1.24
2.22
Index component value
0.48
0.45
0.43
1.24
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
1.46
P
(mW)
133.12
93.42
106.85
P
1×1
(mW)
31.51
34.87
z
s
(cm)
1.65
z
b
(cm)
2.32
z
MI
(cm)
0.50
z
pii
,
α
(cm)
0.50
f
awf
(MHz)
2.53
3.39
2.77
2.76
Other
Informatio
n
prr
(Hz)
699.00
srr
(Hz)
41.00
n
pps
/
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
92.57
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
96.03
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
130.44
p
r
at
z
pii
(MPa)
1.21
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
14.8cm
4.6cm
14.8cm
14.8cm
B Focus Position
6.0cm
4.0cm
6.0cm
2.0cm
SV Position
6.0cm
4.0cm
6.0cm
2.0cm
B Working Frequency
2
3.5
2
2
PW Working Frequency
2.5
3
2.5
2.5
B PRF
3499
8394
1137
1137
PW PRF
699
699
3295
3295
PW SV
0.5mm
0.5mm
0.5mm
0.5mm
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...