Image Optimization 5-3
In real-time scanning of all image modes, press <B> on the control panel to return to B mode.
5.4.2
B Mode Parameters
In B mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
Display
F
D
G
FR
IP
DR
Parameter Frequency Depth
Gain
Frame Rate
B IP
B Dynamic
Range
5.4.3
B Mode Image Optimization
Gain
Description
To adjust the gain of the whole receiving information in B mode. The real-time
gain value is displayed in the image parameter area in the upper left corner of
the screen.
Operation
Rotate the <Gain/iTouch> knob clockwise to increase the gain, and
anticlockwise to decrease.
The adjusting range is 0-100.
Effects
Increasing the gain will brighten the image and you can see more received
signals. However, noise may also be increased.
Depth
Description
This function is used to adjust the display depth of sampling, the real-time
value of which is displayed on the image parameter area in the upper left
corner of the screen.
Operation
Use the <Depth/Zoom> knob to adjust the depth.
The adjustable depth values vary depending upon the probe types.
Effects
Increase the depth to see tissue in deeper locations, while decrease the depth
to see tissue in shallower locations.
Impacts
Depth increase will cause a decrease in the frame rate.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...