Acoustic Output Reporting Table (60601-2-37) C-73
Transducer Model: 75LT38EA
Imaging Mode: B-mode/Smart3D/iScape/Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.91
0.12
0.32
0.32
Index component value
0.12
0.12
0.32
0.12
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
2.08
P
(mW)
18.00
18.00
18.00
P
1×1
(mW)
4.69
4.69
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.46
z
pii
,
α
(cm)
1.46
f
awf
(MHz)
5.28
5.50
5.33
5.39
Other
Informatio
n
prr
(Hz)
8,992.0
0
srr
(Hz)
36.00
n
pps
1.33
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
173.52
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
4.34
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
7.40
p
r
at
z
pii
(MPa)
2.72
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
6.5cm
12.9cm
12.9cm
12.9cm
Focus position
1.0cm
10.0cm
10.0cm
10.0cm
Working Frequency
H10.0
H10.0
H10.0
H10.0
PRF
8992
5094
5094
5094
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...