A-4 Electrical Safety Inspection
LIMITS
For BF
applied parts,
100μA in Normal Condition.
500μA in Single Fault Condition.
A.8
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage,
through a limiting resistance, to selected applied part terminals. Current measurements are
then taken between the selected applied part and earth. Measurements are taken with the
test voltage (110% of mains) to applied parts in the normal and reverse polarity conditions.
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity.
LIMITS
For BF
applied parts: 5000
μA.
A.9
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and
the remaining Applied Part connectors. All measurements may have a true RMS only
response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition);
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF
applied parts,
100μA in Normal Condition.
5
00μA in Single Fault Condition.
Note:
Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
Follow the instructions of the analyzer manufacturer.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...