C-54 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65EL60EA
Imaging Mode: PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.14
0.31
2.79
1.08
Index component value
0.25
0.31
0.95
2.79
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
2.55
P
(mW)
25.75
28.85
29.96
P
1×1
(mW)
10.64
28.85
z
s
(cm)
2.00
z
b
(cm)
1.27
z
MI
(cm)
1.27
z
pii
,
α
(cm)
1.27
f
awf
(MHz)
4.96
4.93
5.02
5.01
Other
Informatio
n
prr
(Hz)
699.00
srr
(Hz)
/
n
pps
/
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
277.85
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
146.63
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
235.44
p
r
at
z
pii
(MPa)
2.81
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
6.5cm
12.9cm
6.5cm
6.5cm
SV Position
1.0cm
10.0cm
1.0cm
0.5cm
Working Frequency
5
5
5
5
PRF
699
6300
23995
23995
SV
0.5mm
0.5mm
0.5mm
0.5mm
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...