1-10 Safety Precautions
8.
Please use the disinfection or sterilization solution that
recommended in this operator’s manual, otherwise Mindray
will not be liable for damage caused by other solutions. If
you have any questions, please contact Mindray Customer
Service Department or sales representative.
9.
The probe sheath contains natural rubber that can cause
allergic reactions in some individuals.
10. Do not use pre-lubricated condoms as a sheath. Lubricant
may not be compatible with the transducer material and
damage may result.
11. Transducer damage may be caused by inappropriate gel,
detergent or cleanser:
Do not soak or saturate transducers with solutions
containing alcohol, bleach, ammonium chloride compounds,
acetone or formaldehyde.
Avoid contact with solutions or coupling gels containing
mineral oil or lanolin.
NOTE:
Read the following precautions to prevent the probe from malfunction:
Clean and disinfect the transducer before and after each examination.
After the examination, wipe off the ultrasound gel thoroughly.
Otherwise, the ultrasound gel may solidify and the image quality would
be degraded.
After an examination, wipe the ultrasound gel off thoroughly. Failure to
do so may result in the ultrasound gel solidifying and the image quality
being degraded.
To prevent the probe from being damaged, do not use it where it will be
exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
Repeated disinfection will eventually damage the probe. Check the probe's
performance periodically.
NOTE:
1. The following definition of the WEEE label applies to EU member
states only: The use of this symbol indicates that this system
should not be treated as household waste. By ensuring that this
system is disposed of correctly, you will help prevent bringing
potential negative consequences to the environment and human
health. For more detailed information with regard to returning and
recycling this system, please consult the distributor from whom
you purchased the system.
2. For system products, this label may be attached to the main unit only.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...