Probes and Biopsy 12-17
12.2.1 Basic Procedures for Biopsy Guiding
1. Select the right needle-guided bracket and needle, install them properly.
Before biopsy guiding, you can preset bracket model, biopsy display and guide line dot
type. For details, please refer to “12.2.3 Biopsy Preset”.
2.
Verify the biopsy guide line. For details, please refer to “12.2.5 Biopsy Menu”.
3. Press <Biopsy> on the control panel.
If the current probe has no matched bracket; or, the image is frozen and the guide
line is hidden before image frozen, then you can’t enter Biopsy menu.
Before entering Biopsy menu, the system will prompt you “Please verify guidelines
before biopsy”.
4.
Set [Display] as “On” to display the guide line, select the bracket and guide line
according to actual situation. For
details, please refer to “12.2.5 Biopsy Menu”.
5. On the Biopsy menu, click [Verify] to enter the Verify menu to verify the guide line. After
the verification, click [Save] on the menu to save the parameter setting. And then click
[Exit] to return to Biopsy menu.
Note:
If you switch the probe or needle-guided bracket during the biopsy, please verify the
guide line again.
When exit the Biopsy Verify menu without saving the setting, system will confirm with
you “Data have changed. Do you want to save the changes?”, Click [OK] to save the
setting and return to Biopsy menu. Click [OK] to save the setting and return to biopsy
menu.
6. Scan to locate the target. Center the target in the electronic guidezone path.
7. Direct the needle into the area of interest for specimen.
8. After extraction of the biopsy sample is completed, gently remove the probe from the
body. To exit Biopsy menu: press <Esc>; or, navigate the cursor onto the Other menu.
9. Disassemble the items and properly dispose of these items as required.
DANGER:
Fail to match the guidezone displayed to the guide may cause the
needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
angle set on the guide, otherwise the needle will not follow the
displayed guidezone which could result in repeated biopsies or
animal injury.
12.2.2 Needle-guided Brackets
A needle-guided bracket is available for purchase as an optional accessory; it is used in
combination with this transducer. Part of the probes have matched needle-guided bracket
and needles. To order needle-guided brackets, contact MINDRAY Customer Service
Department or sales representative.
For biopsy or treatment, ultrasound-guided biopsy procedures can be performed using this
transducer in combination with a needle-guided bracket (optional accessory) and a biopsy
needle (provided by the user).
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...