A-2 Electrical Safety Inspection
A.2
Device Enclosure and Accessories
A.2.1 Visual Inspection
Test Item
Acceptance Criteria
The enclosure and accessories
No physical damage to the enclosure and
accessories.
No physical damage to meters, switches,
connectors, etc.
No residue of fluid spillage (e.g., water,
coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs,
dials, terminals, etc.).
A.2.2 Contextual Inspection
Test Item
Acceptance Criteria
The enclosure and accessories
No unusual noises (e.g., a rattle inside the
case).
No unusual smells (e.g., burning or smoky
smells, particularly from ventilation holes).
No taped notes that may suggest device
deficiencies or operator concerns.
A.3
Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
Main unit label
Integrated warning labels
A.4
Protective Earth Resistance
a. Plug the probes of the analyzer into the
device’s protective earth terminal and protective
earth terminal of the AC power cord.
b. Test the earth resistance with a current of 25 A.
c. Verify the resistance is less than limits.
LIMITS
ALL COUNTRIES R = 0.2
Ω Maximum
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...