12-12 Probes and Biopsy
Before safety and performance is affected, probe 75LT38EA can be sterilized by Cidex
Activated Glutaraldehyde
Solution for at least 217 times (10 hours for one time).
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the sterilization solution, method of sterilization and dilution and
cautions during use.
Do not soak the probe connector or the cable near it into water or any solution.
Follow local regulations when selecting and using the sterilization solution.
3. Rinse the probe with plenty of sterile water (about 2 gallons) for at least 1 minute to
remove all chemical residues on it. Or, follow the rinsing method recommended by the
sterilization solution manufacturer to rinse the probe.
4. Wipe off the water on the probe with sterile cloth or gauze after rinsing it. Do not dry the
probe by heating.
Immerse the probe 75LT38EA in the solution
NOTE:
Repeated disinfection may degrade the performance and safety of the probe.
12.1.6 Environment
Working condition
Use the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
35C50EA
0-40
℃
30%-85%RH
(
no
condensation
)
700-1060hPa
65C15EA
0
℃
-40
℃
30%~85%RH
(
no
condensation
)
700 hPa ~ 1060 hPa
65EB10EA
0
℃~
40
℃
30%
~
90%RH
(
no
condensation
)
700hPa
~
1060hPa
75L38EA
0
℃
-40
℃
30%~85%RH
(
no
condensation
)
700 hPa ~ 1060 hPa
75L53EA
0
℃~
40
℃
30%
~
85%RH
(
no
condensation
)
700hPa
~
1060hPa
10L24EA
0
℃~
40
℃
30%
~
90%RH
(
no
condensation
)
700hPa
~
1060hPa
35C20EA
0
℃
-40
℃
30%~85%RH
(
no
condensation
)
700 hPa ~ 1060 hPa
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...