Multitron - Operating Manual
Options
20 August 2020
Page
51
of
158
4.6
Removable Shelf
The removable shelf (approx. 865 mm x 490 mm
)
allows static incu-
bation in climatic conditions almost identical to the cultivation flasks.
The removable shelf can be inserted at three different heights into
the rails mounted on the side walls in the incubation chamber.
Lower position
If the removable shelf is in the lower position, shake flasks with a vol-
ume of up to 250 millilitre (maximum height: 150 mm) can be placed
on the tray.
Middle position
If the removable shelf is in the middle position, shake flasks with a
volume of up to 1 litre (maximum height: 220 mm) can be placed on
the tray.
Upper position
If the removable shelf is in the upper position, shake flasks with a vol-
ume of up to 3 litres (maximum height: 320 mm) can be placed on
the tray.
ATTENTION
The inserted removable shelf can change the climate conditions in
the incubation chamber and lead to higher tolerances as it hinders
air circulation. To prevent damage to cultures, the climate above
and beneath the removable shelf should be checked frequently
and, if necessary, monitored using mobile sensors.