Multitron - Operating Manual
Operation
20 August 2020
Page
77
of
158
The shaker drive and all other parameters will start up again auto-
matically as soon as the door is closed.
INFORMATION
If you wish to stop the shaker drive slowly, e.g. to avoid culture liq-
uid from spilling out, you have to set the
Rotation speed
parame-
ter to
OFF
prior to opening the door. In this case, the shaker drive
stops slowly, with the brake force set in the settings menu (see
chapter 7.7.4 "Setting the Brake Force for Stopping the Table",
page 112).
ATTENTION
The door can only be opened fully when there is no load on the
half-open door. If the door is pushed down violently, this can
damage the door mechanism.
Door alarm
To prevent the door from being left open unintentionally, the device
has a door alarm (beeping). This is triggered if the door is left open
for longer than the set time. If no parameter was active prior to
opening the door, the door alarm is not triggered. The time until the
door alarm is triggered can be set in the settings menu (1 to
10 minutes). If necessary, the door alarm can also be deactivated
completely.
7.2.2
Inserting and Removing the Tray
The tray is affixed to the table in different ways depending on the
throw.
Device with 25 mm or 50 mm throw:
The device is fitted with a tray lock to hold the tray in place on the ta-
ble. To insert and affix the tray, manually move it into the incubation
chamber all the way to the limit stop. When you close the door, the
tray is lowered into the locking cones (2) and thus affixed to the ta-
ble.
When you open the door completely, the ejection levers (1) under the
table lift the tray out of the table’s locking cones (2). The tray is now
free and can be pulled out of the incubation chamber by hand.
2
1