Faults in conjunction with the
temperature
parameter
Fault
Temperature too high compared to setpoint.
Possible cause
Measures
Personnel
Fault in temperature measur-
ing.
Check if the Pt100 sensor works. In case of a de-
fect, contact your local INFORS HT representative.
Operator
Internal or external cooling unit
is not switched on.
Switch on cooling.
Operator
No cooling available and differ-
ence between setpoint value
and ambient temperature too
small.
Increase the setpoint value or equip the device
with a cooling.
Operator
Fault
The temperature does not reach the desired setpoint.
Possible cause
Measures
Personnel
The heating is not switched on. Activate the
temperature
parameter.
Operator
The setpoint is set too low.
Raise setpoint.
Operator
The ambient temperature does
not meet the requirements.
Check the ambient temperature and correct it if
necessary (see chapter 6.1, page 72
Operator
Door not closed completely.
Close door completely.
Operator
The heating fans do not work. Contact your local INFORS HT representative.
INFORS HT service tech-
nician or licenced dealer
INFORMATION
The re-measuring of the temperature only provides reliable infor-
mation when calibrated measuring tools are used and only if they
are used at points specified by INFORS HT. A measurement at un-
defined positions in the casing will not produce any usable data.
For information about temperature measuring, contact your local
INFORS HT representative or request a quote for calibrating the pa-
rameters.