7.5
Timer Function
7.5.1
Overview
The timer function allows you to programme defined sequences (e.g.
changing parameter setpoints after a defined time period or delaying
the start of the cultivation process). The timer can be programmed in
two different modes:
One-time change from phase 1 to phase 2
Cyclical change between phase 1 and phase 2
INFORMATION
More complex sequences and procedures (e.g. daylight curves) can
be realised with the bioprocess platform software eve
®
.
Signs regarding the timer function
If the timer function is active, the following signs appear:
The sign
Timer
appears as soon as the timer function is active or if
the timer can be programmed.
The signs
Time 1
and
Time 2
indicate, that the duration of phase 1
(
Time 1
) respectively phase 2 (
Time 2
) can be entered.
The signs
Phase 1
and
Phase 2
indicate which phase is active or for
which phase the parameter setpoints can be entered:
The sign is blinking: The phase is active.
The sign is lit up: The parameter setpoints of the corresponding
phase can be entered.
If the cyclic change between phase 1 and phase 2 is active, the sign
cycle
lights up.