Multitron - Operating Manual
Operation
20 August 2020
Page
101
of
158
Checking the timer for the second phase
7.
Tap the
TIMER
key to get to the time settings for the second
phase (
Time 2
).
On the lower alphanumeric display appears the time entered last
or 00:00 if no time has been entered.
In the separate timer display area the signs
Timer
and
Phase 2
are lit up.
8.
Make sure the second timer is turned off. If not, use the
ON/OFF
key in the lower setting key area to turn the timer
(
Time 2
) off.
Entering the setpoints for the second phase and switching on
the parameters
After making sure, that the second timer (
Time 2
) is switched off, the
different parameters for phase 2 can be selected using the
Select
key
and afterwards be switched on.
INFORMATION
If the timer should be used to stop the cultivation process after a
defined time, all parameters for phase 2 can be switched off. This
causes the parameters to switch off after the set time for phase 1
has elapsed and therefore to stop the cultivation process.
9.
Tap the
Select
key to select the desired parameters.
In both display areas the first integral pair of parameters (
Temp
and
RPM
) appears with the last setpoint values entered or the
message
OFF
.
The sign
Setpoint
indicates, that the setpoints can be entered.