Multitron - Operating Manual
Operation
20 August 2020
Page
105
of
158
7.5.4
Changing Timer Settings during Activated Timer Function
If it is required to change the duration of the time intervals, both tim-
ers can be reprogrammed. To change the setting of the timers pro-
ceed as follows:
Procedure
1.
Tap the
TIMER
key to select the timer function.
2.
Tap again the
TIMER
key to enter the timer programming
mode.
On the upper alphanumeric display the setpoint value last en-
tered for the time interval of the first phase (
Time 1)
is shown.
The sign
Setpoint
indicates, that the time can be entered.
3.
If required, tap the
Plus
or the
Minus
key in the upper setting
key area to set the time for the first phase (
Time 1
).
INFORMATION
Do not tap the
ON/OFF
key after changing the timer setting, as
this would turn off the timer immediately.
If this happens in the cycle mode with the timer for the first phase,
also the timer for the second phase is deactivated. And if acci-
dentally the second timer is turned off, it cannot be activated
again.
4.
Tap the
TIMER
key to get to the time settings for the second
phase (
Time 2
).
On the lower alphanumeric display the setpoint value last en-
tered for the time interval of the second phase (
Time 2)
is
shown. The sign
Setpoint
indicates, that the time can be en-
tered.
5.
If required, tap the
Plus
or the
Minus
key in the lower setting
key area to set the time for the second phase (
Time 2
).
6.
Tap the
TIMER
key again to quit the programming mode of the
timer.
On the alphanumeric display the remaining time of the active
phase and the message
run
appear.
7.
Tap the
Select
key return to the normal operating mode of the
display.