Multitron - Operating Manual
Operation
Page
100
of
158
20 August 2020
4.
Tap the
ON/OFF
key in the upper setting key area to activate
the timer (
Time 1
).
On the upper alphanumeric display the message
on
appears for
a very short time and then the time entered is shown
In the separate timer display area the sign
Timer
is lit up and the
sign
Phase 1
lights up for the first time.
INFORMATION
It is not possible to set the parameters for the first phase as long as
the timer for the first phase has not been activated.
Entering the setpoints for the first phase and switching on the
parameters
As soon as the timer for the first phase (
Time 1
) has been switched
on, the different parameters for phase 1 can be selected using the
Select
key and afterwards be switched on.
5.
Tap the
Select
key to select the desired parameters.
In both display areas the first integral pair of parameters (
Temp
and
RPM
) appears with the last setpoint values entered or the
message
OFF
.
The sign
Setpoint
indicates, that the setpoints can be entered.
INFORMATION
In this phase of the programming process the sign
Setpoint
re-
mains lit up and does not disappear after 10 seconds.
6.
Tap the
Plus
or
Minus
key in the corresponding setting key area
to set the desired setpoint values and switch the parameters on
using the
ON/OFF
key.
INFORMATION
With the
Select
key the parameters and the time entered for the
first phase can be selected and changed over and over again.