Placing the device under a table
When placing the device under a table, observe the following:
You must leave a ventilation gap of at least 100 mm between
the device and rear wall of the table to ensure heat produced
by the device can escape.
The minimum distance between the device and the underside
of the table top is 100 mm.
When operating the device with installed cooling, heat can ac-
cumulate under a closed laboratory bench. This can result in
the desired temperature in the incubation chamber not being
reached. In addition, heat accumulation increases power con-
sumption, as more cooling is required. Hence, we recommend
only installing the device under a laboratory bench with air
vents.
ATTENTION
Failure to comply with the required minimum distances can result
in components of the device overheating and the device being
damaged. Therefore, please note:
–
Never cover the air vents on the right side or those on the
rear of the device.
–
Never position the device directly on a wall.