Multitron - Operating Manual
Operation
20 August 2020
Page
103
of
158
13.
Tap the
Select
key to quit the timer programming mode.
If the
Select
key is not used, the display changes to the normal
operating mode automatically after 60 seconds.
The displays now show the current values of the integral param-
eters
temperature
and
rotation speed
. If other parameters are
installed, the display shows the current values of each pair of pa-
rameters for 20 seconds.
The cultivating process now runs with the parameter setpoint values
for the first phase, until the time entered for the first timer (
Time 1
)
has elapsed.
After the time for the first phase has lapsed, the message
End
ap-
pears on the lower alphanumeric display. This message alternates
every 20 seconds with the current values of the parameters for the
second phase. In addition, there is an acoustic alarm every 60 sec-
onds.
The cultivating process now runs with the setpoint values of the sec-
ond phase until the device is stopped manually by turning the param-
eters off. If the alarm is confirmed by tapping the
TIMER
key, the
acoustic alarm stops, the message
End
does not appear anymore.
INFORMATION
The timer function can at any time be deactivated separately or by
turning off the first timer. The cultivating process then continues
with the last active parameter setpoint values.