Multitron - Operating Manual
Operation
Page
106
of
158
20 August 2020
7.5.5
Changing Parameter Setpoint Values during Activated Timer Function
Especially during long cultivating processes in the cycle mode it might
be required to change the parameter settings after a certain time.
Parameters of the active phase
or (in the cycle mode)
Which phase is active is indicated by the blinking of the symbol
Phase
1
or
Phase 2
in the separate timer display area.
To change the setpoint values of the parameters of the active phase
proceed as follows:
Procedure
1.
Select the desired pair of parameters by tapping the
Select
key
– if necessary several times.
In the two alphanumeric display areas the setpoint values of the
selected pair of parameters are shown.
2.
Tap the
SET
key in the upper or lower operation key area.
The sign
Setpoint
indicates, that a setpoint can be set.
3.
Tap the
Plus
or
Minus
key in the according setting key area to
set the new setpoint value.
The new setpoint value for the parameter appears on the ac-
cording alphanumeric display.
4.
Tap the
Select
key to quit the parameter setting mode.
The display returns to the normal operating mode and shows the cur-
rent values of the parameters.