Multitron - Operating Manual
Rectifying Faults
Page
120
of
158
20 August 2020
8.2
Faults and Error Messages
8.2.1
Error Messages Explained
The faults listed here, the causes of which are described using error
messages, can generally not be resolved by users. One of the manu-
facturer’s service technicians needs to be consulted if these messages
appear.
In addition to the warning indicator and the
ERROR
sign, messages
that are abbreviated appear at the top and bottom alphanumeric dis-
plays. These messages have the following meanings:
Top
Bottom
General faults
Open
The device door is open. When the door is opened, the device is au-
tomatically stopped using the greatest possible breaking effect.
Top
Bottom
In conjunction with the
temperature
parameter
(
Temp
symbol lights up)
Error 1, Temperature High
A temperature above 100 °C is measured.
Error 1, Temperature Low
A temperature below 0 °C is measured.
Error 1, Sensor
The Pt100 sensor is not returning any readings.
Error 1, Fan 1 - 4
One of the fans is blocked or defective.