4.2.3
Operating the Direct Steam Humidification
The direct steam humidification is operated using the
Humidity
pa-
rameter. The bottom of the alphanumeric display shows the value in
%
(relative humidity) with the symbol
Humid
.
After switching on the parameter, it takes approx. 5 minutes for the
steam generator to heat up. During the heating time, the current ac-
tual humidity value and the message
Strt
are displayed alternately.
To prevent condensation on the walls in the incubation chamber, the
direct steam humidification is only switched on when the set temper-
ature setpoint (± 1 °C) is reached and stable for at least 1 minute.
During the time until the temperature is stable, the message
nA
is
displayed.
For a detailed description of how to set and activate the parameters,
see chapter 7.4 "Adjusting, Activating and Deactivating Parameters",
page 90.
4.2.4
Specifications and Technical Data
For a detailed description of the technical data and possible humidity
values, see chapter 12.5.3 "Direct Steam Humidification", page 148.