3.2.2
Standard Function Temperature Control (Heating)
1
Finned heating element and radial fans (in back
wall, not visible)
2 Pt100 temperature sensor
The temperature in the incubation chamber is regulated using a
Pt100 temperature sensor with a PID controller. Four radial fans en-
sure constant air circulation and keep the temperature distribution in
the incubation chamber as constant and gradient-free as possible. A
finned heating element heats the air in the incubation chamber. In
order to reach temperatures below the ambient temperature, the de-
vice can be fitted with an optional cooling system.
INFORMATION
The temperature sensor serves exclusively to measure the tempera-
ture of the air in the incubation chamber. The temperature sensor
must not be removed from the holder and immersed in the liquid
in the cultivation vessel.
Operation
The
Temperature
parameter is operated using the operating panel.
Beside the actual value, the symbol
Temp
and unit
°C
appear on the
display.
For an exact description of how to set the parameters, see chapter
7.4 "Adjusting, Activating and Deactivating Parameters", page 90.
Technical data and setpoints
For a detailed description of the technical data and possible maxi-
mum or minimum temperatures, see chapter 12.4.2 "Temperature
Parameter (Heating and Ventilation)", page 145.