Specifications for the connection / water quality
Description
Value Unit
Connection (hose Ø)
6 to 7 mm
Input pressure
0 to 0.3 bar
Water hardness
< 0.01, CaCO
3
equivalent
mmol/L
Water conductivity
min. 5 / max. 20
μS/cm
Dissolved solids
< 10 mg/L
ATTENTION
Using tap water can quickly lead to limescale in the vaporiser of
the humidification unit, which would affect correct functioning.
Ultra pure water (WFI, water for injection purposes) must not be
used either as this is highly corrosive and thus damages the device.
Adding detergents, disinfectants or similar chemicals to the water
for the direct steam humidification can damage the device. Only
use water without additions, in accordance to the specification.
Specifications for the humidity sensor
Description
Value Unit
Sensor type
EE071 HCT01-00D
Measurement range
20 to 90 %
Response time
< 300 ms
Temperature range for use
- 40 to 80 °C
12.5.4
Pressure Elimination Unit for Direct Steam Humidification
Dimensions
Description
Value Unit
Depth
185 mm
Height
337 mm
INFORMATION
For detailed information on the external dimensions, refer to the
dimension drawings (see chapter 4.3.4 "Dimensions", page 47).