Multitron - Operating Manual
Rectifying Faults
Page
122
of
158
20 August 2020
Top
Bottom
In combination with the
CO
2
parameter
(
CO
2
symbol lights up)
Error, Sensor
The CO
2
sensor is not returning any readings (CO
2
sensor not con-
nected or defective).
Error, Control
The actual CO
2
value does not rise (CO
2
supply interrupted, pressure
too low or door not closed correctly).
Top
Bottom
In conjunction with the control
Error, Program
Due to a new program version, problems with EEProm or problems
with one of the internal data buses, EEProm was initialised.
After this fault, it might be necessary to recalibrate temperature
measuring again.
Error, Lock
The message appears when the key pad is locked with a PIN. If the
message appears although the PIN function is not active, the key
pad cannot be initialised. If the problem is caused by one of the in-
ternal data buses, the mainboard or HMI unit must be replaced.