7.7.2
Setting the Upper Limit for the Temperature
Top display:
OPt
,
Temp
,
Setpoint
and
High
The maximum temperature that can be entered as a setpoint can be
restricted. To do so, tap the
SET
key (the sign
Setpoint
appears) and
then use the
Plus
or
Minus
key to set the desired maximum temper-
ature.
Default value: 65 °C
7.7.3
Setting the Lower Limit for the Temperature
Top display:
OPt
,
Temp
,
Setpoint
and
Low
The minimum temperature that can be entered as a setpoint can be
restricted. To do so, tap the
SET
key (the sign
Setpoint
appears) and
then use the
Plus
or
Minus
key to set the desired minimum temper-
ature.
Default value: 4.0 °C
7.7.4
Setting the Brake Force for Stopping the Table
…
Top display:
OPt
Bottom display:
br. 0-3
The brake force with which shaking operation is stopped when
switching off the parameter
rotation speed
can be set in four levels.
To do so, tap the
SET
key (the sign
Setpoint
appears) and then use
the
Plus
or
Minus
key to set the desired brake force.
The levels
br. 0
to
br. 3
have the following meaning:
At brake force level
0
the table is stopped by reducing the
speed very slowly.
At brake force level
1
the table is also controlled but stopped a
little less slowly.
At brake force level
2
a passive brake is activated by switching
off the motor (factory setting).
At brake force level
3
the motor is short-circuited to stop the
table as quickly as possible.