Stacked devices
With devices stacked in three, very low loads (< 10 kg per unit) and
high rotation speeds (> 300 min
-1
) at the same time can result in vi-
brations occur. In this case it is recommended to place additional cul-
tivation vessels filled with water on the tray. This ensures a smooth
run. With a load of more than 10 kg per unit, no significant vibra-
tions occur and the device can be operated over the entire rotation
speed range without any limitations.
WARNING
If the loading mass of the table is too high or too low or the load
is distributed unevenly, high rotation speeds can cause vibrations.
This can result in the device moving uncontrollably.
–
Do not operate the device without tray and loading.
–
In case of strong vibrations, reduce the rotation speed and