7.9
Switching Off the Device
ATTENTION
Parameters, that have not been switched off prior to switching off
the device, are automatically activated when the device is switched
on. This can lead to damages to the device and cultivation vessels.
Always switch off all parameters before switching off the device.
INFORMATION
The parameter setpoints remain stored for approximately one
month.
To switch off the device, proceed as follows:
Procedure
1.
Switch off all parameters. Make sure that, in addition to the
ro-
tation speed
, parameters that are not visible, such as
tempera-
ture
,
humidity
or
CO
2
, are also switched off.
2.
Switch off the device by pressing the power switch.
3.
If you do not intend to use the device for a long time, discon-
nect the device from the power supply.
7.10
Behaviour in Case of Interrupted Power Supply
If the power supply to the device is interrupted during a running cul-
tivation process (e.g. by flicking the power switch or in case of a
power failure), all parameters and timer setpoints as well as the resid-
ual time of the last active timer phase are stored.
If power supply is restored, the device restarts automatically with the
last stored setpoints. If a timer was active prior to the interruption to
power supply, the device restarts with the residual time of the last
active phase and the setpoints stored for this phase.
As a warning, the word
RESTARTED
and the warning icon flash on
the display field with the warning and alarm messages.
The
RESTARTED
message can be confirmed by tapping any key and
then disappears.