Multitron - Operating Manual
Rectifying Faults
Page
124
of
158
20 August 2020
In conjunction with the
rotation speed
parameter
Fault
The table does not move.
Possible cause
Measures
Personnel
The table is blocked by foreign
objects.
Remove foreign objects (see chapter 9.2, page
129).
Operator
The drive belt has torn.
Contact your local INFORS HT representative.
INFORS HT service tech-
nician or licenced dealer
The motor is defective.
Contact your local INFORS HT representative.
INFORS HT service tech-
nician or licenced dealer
The (open-loop) control of the
drive is defective.
Contact your local INFORS HT representative.
INFORS HT service tech-
nician or licenced dealer
Fault
Strong vibrations occur.
Possible cause
Measures
Personnel
The load is distributed une-
venly.
Load the tray in the centre. If possible, do not
place any heavy weights in the corners of the tray.
Operator
The rotation speed is too high. Reduce the rotation speed.
Operator
The device is not standing
straight.
Level the table or the device (adjustable foot on
the base).
Operator
The substructure is too weak.
Place the device on a stable floor.
Operator
Fault
The shaker drive accelerates too quickly or runs too fast.
Possible cause
Measures
Personnel
Fault in measuring the rotation
speed.
Switch the device off and then on again.
Operator
The motor control is defective. Contact your local INFORS HT representative.
INFORS HT service tech-
nician or licenced dealer